Goniophotometer for lamps and small luminaire

      The goniophotometer type RIGO 801 utilizes a new image-resolving CCD measuring technique for determining
ray data and luminous intensity distributions. The correct determination of the luminous intensity distributions (LVK) of lamps and luminaries is performed far within their photometric limiting distance on the basis of image resolved measurements of luminous intensity distributions. A CCD-camera is moved by a goniometer around the measuring object at rest on a spherical surface, with the
radius of this sphere being fixed only by the field angle of the camera. Thus, the goniometer can also be installed in small laboratories

Feature

  • Correct determination of the luminous intensity distributions (LID) of lamps and luminaries far within their photometric near-field range
  • Output of lid data in all common file formats (EULUM-DAT, TM14, IES, Calculux, ...)
  • Capture of ray data
  • Saving in various standard formats (ASAP, Speos, Zemax, LightTools, LucidShape, TracePro)
  • Small dimensions of the set-up
  • Measuring in normal position, no movement of the measuring object photometric database(LumCAT)

Application

  • LIDs, spectral characterization; object assumed to be a point
  • luminance distributions
  • 4D-luminance distribution, ray distribution in the space
  • Efficiency ratio, power input, temperature behavior

 

Specification

Size of measuring object 20 x 20 x 20 mm³ ~ 300 x 300 x 300 mm³
Space required 1300 x 1300 x 1900 mm³
Movement The measuring camera and the illuminance meter are moved on a sphere around the lamp (two independent axes arranged vertically
to each other (θ, Φ)).
Travel path Φ = -2° ~ 362°, θ = 0° ~ 360°
Measuring position of the luminaire Normal position, no movement of the measuring object. The whole goniometer can be swivelled, which permits different measuring positions to be realized.
Measuring width 270 mm
Positioning accuracy Φ < 0.02°, θ < 0.05°
Repetitive accuracy Φ < 0.01°, θ < 0.02°

Option

AC voltage stabilizer SX1H-X20S (Statron AG)
DC laboratory power supply type 500-52, 62N52 RU 25P (GMC-Instruments)
Digital wattmeter type WT200 (Yokogawa)
Control panel
Standard lamp for luminous flux (LMT)
Measuring lampholder for standard lamp