Goniophotometer for
luminaire |
The goniophotometer type RIGO
801 utilizes a new image-resolving CCD measuring technique for
determining
ray data and luminous intensity distributions. The correct
determination of the luminous intensity distributions (LVK) of lamps
and luminaries is performed far within their photometric limiting
distance on the basis of image-resolved measurements of luminous
intensity distributions. A CCD-camera is moved by a goniometer
around the measuring object at rest on a spherical surface, with the
radius of this sphere being fixed only by the field angle of the
camera. Thus, the goniometer can also be installed in small
laboratories |
|
Feature |
-
Correct determination of the
luminous intensity distributions
(LID) of lamps and luminaries far
within their photometric near-field
range
-
Output of lid data in all common
file formats (EULUM-DAT, TM14, IES,
Calculux, ...)
-
Capture of ray data
-
Saving in various standard formats
(ASAP, Speos, Zemax, LightTools,
LucidShape, TracePro)
-
Small dimensions of the set-up
-
Measuring in normal position, no
movement of the measuring object
photometric database(LumCAT)
|
|
Application |
- LIDs, spectral characterization;
object assumed to be a point
- luminance distributions
- 4D-luminance distribution, ray
distribution in the space
-
Efficiency ratio, power input,
temperature behavior
|
|
Specification
|
|
Size of measuring object |
≤ 2000 x 2000 x 2000 mm³ |
| Space required |
≤ 4000 x 4000 x 4900 mm³ |
|
Movement |
The measuring camera and the illuminance meter are moved
on a sphere around the lamp (two independent axes
arranged vertically to each other (θ, Φ)). mounted to
a fixed upper point. |
|
Travel path |
Φ = -2° ~ 362°, θ = 0° ~ 360° |
|
Measuring position of the luminaire |
Normal position, no movement of the measuring object;
lamp suspension possible at the top and at the bottom. |
| Measuring width |
500 ~ 2000 mm |
|
Positioning accuracy |
Φ < 0.02°, θ < 0.05° |
|
Repetitive accuracy |
Φ < 0.01°, θ < 0.02° |
|
Option |
|
AC voltage stabilizer SX1H-X20S (Statron AG) |
|
DC laboratory power supply type 500-52, 62N52 RU 25P
(GMC-Instruments) |
|
Digital wattmeter type WT200 (Yokogawa) |
|
Control panel |
|
Standard lamp for luminous flux (LMT) |
|