Specification
|
| Model No. |
Luminance Meter
LS-100 |
Luminance Meter
LS-110 |
| Type |
SLR
spot luminance meter for measuring light-source and surface
brightness |
| Acceptance Angle |
1° |
1/3° |
| Optical System |
85mm
f/2.8 lens; SLR viewing system; Flare factor less than 1.5% |
| Angle of View |
9° |
| Focusing Distance |
1014mm
(40 in.) to infinity |
| Minimum Measuring
Area |
Ø14.4mm |
Ø4.8mm |
| Receptors |
Silicon
photocell |
| Relative Spectral
Response |
Within
8% (f1') of the CIE spectral luminous efficiency V(λ) |
| Response Time |
FAST:
Sampling time: 0.1sec, Time to display: 0.8 to 1.0sec
SLOW: Sampling time: 0.4sec, Time to display: 1.4 to 1.6sec |
| Luminance Units |
cd/m²
or fL (switchable) |
| Measuring Range |
FAST: 0.001 to
299,900 cd/m² (0.001 to 87,530fL)
SLOW: 0.001 to 49,990 cd/m² (0.001 to 14,590fL) |
FAST: 0.01 to
999,900cd/m²
(0.01 to 291,800fL)
SLOW: 0.01 to 499,900cd/m² (0.01 to 145,900fL) |
| Accuracy |
0.001 to 0.999cd/m²
(or fL):
± 2% ± 2 digits of displayed value
1.000cd/m² (or fL) or greater:
± 2% ± 1 digit of displayed value (Measurement subject:
Illuminant A) |
0.01 to 9.99cd/m² (or
fL):
± 2% ± 2 digits of displayed value
10.00cd/m² (or fL) or greater:
± 2% ± 1 digit of displayed value (Measurement subject:
Illuminant A) |
| Repeatability |
0.001 to 0.999cd/m²
(or fL):
± 0.2% ± 2 digits of displayed value
1.000cd/m² (or fL) or greater:
± 0.2% ± 1 digit of displayed value (Measurement subject:
Illuminant A) |
0.01 to 9.99cd/m² (or
fL):
± 0.2% ± 2 digits of displayed value
10.00cd/m² (or fL) or greater:
± 0.2% ± 1 digit of displayed value (Measurement subject:
Illuminant A) |
| Calibration Mode |
Konica
Minolta standard / User-selected standard (switchable) |
| Color Correction
Factor |
Set by
numerical input; Range: 0.001 to 9.999 |
| Reference Luminance |
1; Set
by measurement or numerical input |
| Measurement Modes |
Luminance; Luminance ratio; Peak luminance or luminance
ratio |
| Display |
External:
4-digit LCD with additional indications
Viewfinder: 4-digit LCD with LED backlight |
| Data Communication |
RS-232C; Baud rate: 4800bps |
| External Control |
Measurement process can be started by external device
connected to data output terminal |
| Power Source |
One 9V
battery; Power can also be supplied by optional Data Printer
DP-10 |
| Operation
Environment Conditions |
Temperature: 0 to 40° C (32 to 104° F); Relative humidity
85% or less (at 35° C / 95° F) with no condensation, Maximum
altitude: 2000m, Installation category: II, Pollution
degree: 2 |
| Storage Temperature
Range |
-20 to
55° C (-4 to 131° F); Relative humidity 85% or less (at 35°
C / 95° F) with no condensation |
| Dimensions (W x H x
D) |
79 x
208 x 150mm |
| Weight |
850g
without battery |
| Standard
Accessories |
Lens
Cap, Eyepiece Cap, ND Eyepiece Filter, 9V Battery, Case |