E-Mail

HomeChroma MeterImageing PhotometerIntegrating SphereGonio Photometer

Spectroradio Meter CS-1000A/CS-1000S/CS-1000T

       The accuracy.CS-1000A, CS-1000S, and CS-1000T perform measurements of physical energy (absolute values) and displays the results at 1nm intervals in the visible range of frequencies between 380nm and 780nm. In addition, they also use these measurement values to calculate luminance, chromaticity, color temperature, dominant wavelength, and other similar factors with a high degree of accuracy.
        These spectroradiometers can be used to perform measurements without the need for additional equipment, and their compact size ensures portability for a wide range of applications; accordingly, any specimen can be measured by simply bringing your CS-1000 to it.
       Windows-compatible data management software is bundled as standard. A PC can be used to conveniently control your CS-1000 or to display, process, and save measurement results. In addition, this product is also capable of measuring object colors.
Feature
  • Wide measuring range from low luminance to high.
  • Performance comparable to many spectroradio meters.
  • Compact and lightweight. Battery power is also possible.
  • Selectable measuring angle
  • Data management software CS-S10w standard edition and optional professional edition are provided.
  • Additional Functions

Application

  • Measurement of spectral data, luminance, chromaticity, and correlated color temperature for CRTs, LCDs, OLEDs, and other display devices.
  • Measurement of spectral data, luminance, chromaticity, and correlated color temperature for lamps and other light sources.
  • Reference device for a wide range of luminance and chromaticity measurement tools.
  • Non-contact measurement of object color.

Specification

Model No. CS-1000A CS-1000S CS-1000T
Wavelength range 380 to 780nm
Spectral bandwidth 5nm (half bandwidth)
Wavelength resolution 0.9nm/pixel
Display wavelength bandwidth 1nm
Wavelength precision ±0.3nm (median wavelength: 546.1nm Hg lamp)
Acceptance angle 1° (with standard or macro lens) --- (Standard lens: 1°) 0.14 (Standard lens: 1°)
Display Lvxy, ΔLvxy, Lvu'v', ΔLvu'v', LvTΔuv, Le (Observer switchable between 2° and 10°)
Data memory Measurement data: 30 sets; Target data: 20 sets
Minimum measuring distance 362mm (Standard lens)
94mm (Macro lens)
25mm (Standard lens: 362mm) 254mm
(Standard lens: 362mm)
Minimum measuring area 7.9mm (Standard lens)
1.15mm (Macro lens)
0.45mm
(Standard lens: 7.9mm)
1.2mm
(Standard lens: 7.9mm)
Luminance display range 0.01-80000cd/m² (For Illuminant A)
Accuracy (For Illuminant A, Normal Mode) Luminance: ±2% ±1 digit
Chromaticity:
x: ±0.0015
y: ± 0.001
(Standard or macro lens)
Luminance: ±2.5%, ±1 digit
Chromaticity:
x: ±0.002
y: ±0.0015
(Small measuring area lens)
Luminance: ±2.5%, ±1 digit
Chromaticity:
x: ±0.002
y: ±0.0015
(small measuring angle lens)
Luminance range
Standard lens: 1 to 8000cd/m²
Other lenses: 10 to 80000cd/m²
Repeatability (σ ) (For Illuminant A) Normal Mode Luminance: 0.1%, +1 digit
Chromaticity xy: 0.0002 Luminance range
Standard lens: 1 to 8000cd/m²
Other lenses: 10 to 80000cd/m²
Fast Mode
Luminance: 0.1% +1 digit
Chromaticity xy: 0.0004
Normal Mode
Luminance: 0.1% +1 digit
Chromaticity xy: 0.0003 Luminance range
Standard lens: 0.5 to 1cd/m²
Other lenses: 5 to 10cd/m²
Fast Mode
Luminance: 0.1% +1 digit
Chromaticity xy: 0.0006
Polarization error Less than 5% (400 to 780nm)
Integration time Fast: 40msec to 15sec; Normal: 40msec to 60sec
Power AC 120V, 60Hz or AC 230V 50Hz using AC Adapter AC-A12
Operating temperature/humidity range 5 to 35°C, relative humidity 80% or less with no condensation
Storage temperature/humidity range 0 to 45°C, relative humidity 80% or less with no condensation
Interface RS-232C
Dimensions (main unit) 146 x 148 x 256mm
Weight 4.9kg with standard lens 5.8kg with Small Measuring Area Lens 5.9kg With Small Measuring Angle Lens
Standard accessories Standard Lens
Macro Lens
Standard Lens
Small Measuring Area Lens
Standard Lens
Small Measuring Angle Lens
Data Processing Software CS-S1w; ND Eyepiece Filter (for finder) CS-A1; AC Adapter AC-A12; RS-232C Cable (for IBM PC/AT, 2m, 9-pin) IF-A12
Optional accessories Tripod CS-A3; Panhead CS-A4; White Calibration Plate CS-A5; ND Filter CS-A6 (10%/for Macro Lens); ND filter CS-A7 (1%/for Macro Lens); RS-232C Cables IF-A13 to IF-A15; Hard Case CS-A2 (Not for Small Measuring Area Lens or Small Measuring Angle Lens)