Spectroradio Meter
CS-1000A/CS-1000S/CS-1000T |
The
accuracy.CS-1000A, CS-1000S, and CS-1000T perform measurements of
physical energy (absolute values) and displays the results at 1nm
intervals in the visible range of frequencies between 380nm and
780nm. In addition, they also use these measurement values to
calculate luminance, chromaticity, color temperature, dominant
wavelength, and other similar factors with a high degree of
accuracy.
These spectroradiometers can be used
to perform measurements without the need for additional equipment,
and their compact size ensures portability for a wide range of
applications; accordingly, any specimen can be measured by simply
bringing your CS-1000 to it.
Windows-compatible data
management software is bundled as standard. A PC can be used to
conveniently control your CS-1000 or to display, process, and save
measurement results. In addition, this product is also capable of
measuring object colors. |
|
Feature |
- Wide measuring range from low luminance to high.
- Performance comparable to many spectroradio meters.
- Compact and lightweight. Battery power is also
possible.
- Selectable measuring angle
- Data management software CS-S10w standard edition
and optional professional edition are provided.
- Additional Functions
|
|
Application |
-
Measurement of spectral data, luminance, chromaticity, and
correlated color temperature for CRTs, LCDs, OLEDs, and other
display devices.
-
Measurement of spectral data, luminance, chromaticity, and
correlated color temperature for lamps and other light sources.
-
Reference device for a wide range of luminance and chromaticity
measurement tools.
-
Non-contact measurement of object color.
|
|
Specification |
| Model No. |
CS-1000A |
CS-1000S |
CS-1000T |
| Wavelength range |
380 to
780nm |
| Spectral bandwidth |
5nm
(half bandwidth) |
| Wavelength
resolution |
0.9nm/pixel |
| Display wavelength
bandwidth |
1nm |
| Wavelength
precision |
±0.3nm
(median wavelength: 546.1nm Hg lamp) |
| Acceptance angle |
1° (with standard
or macro lens) |
--- (Standard lens:
1°) |
0.14 (Standard
lens: 1°) |
| Display |
Lvxy,
ΔLvxy, Lvu'v', ΔLvu'v', LvTΔuv, Le (Observer switchable
between 2° and 10°) |
| Data memory |
Measurement data: 30 sets; Target data: 20 sets |
| Minimum measuring
distance |
362mm (Standard lens)
94mm (Macro lens) |
25mm (Standard lens:
362mm) |
254mm
(Standard lens: 362mm) |
| Minimum measuring
area |
7.9mm (Standard lens)
1.15mm (Macro lens) |
0.45mm
(Standard lens: 7.9mm) |
1.2mm
(Standard lens: 7.9mm) |
| Luminance display
range |
0.01-80000cd/m² (For Illuminant A) |
| Accuracy (For
Illuminant A, Normal Mode) |
Luminance: ±2% ±1
digit
Chromaticity:
x: ±0.0015
y: ± 0.001
(Standard or macro lens) |
Luminance: ±2.5%, ±1
digit
Chromaticity:
x: ±0.002
y: ±0.0015
(Small measuring area lens) |
Luminance: ±2.5%, ±1
digit
Chromaticity:
x: ±0.002
y: ±0.0015
(small measuring angle lens) |
Luminance range
Standard lens: 1 to 8000cd/m²
Other lenses: 10 to 80000cd/m² |
| Repeatability (σ )
(For Illuminant A) |
Normal
Mode Luminance: 0.1%, +1 digit
Chromaticity xy: 0.0002 Luminance range
Standard lens: 1 to 8000cd/m²
Other lenses: 10 to 80000cd/m² |
Fast Mode
Luminance: 0.1% +1 digit
Chromaticity xy: 0.0004 |
Normal Mode
Luminance: 0.1% +1 digit
Chromaticity xy: 0.0003 Luminance range
Standard lens: 0.5 to 1cd/m²
Other lenses: 5 to 10cd/m² |
Fast Mode
Luminance: 0.1% +1 digit
Chromaticity xy: 0.0006 |
| Polarization error |
Less
than 5% (400 to 780nm) |
| Integration time |
Fast:
40msec to 15sec; Normal: 40msec to 60sec |
| Power |
AC
120V, 60Hz or AC 230V 50Hz using AC Adapter AC-A12 |
| Operating
temperature/humidity range |
5 to
35°C, relative humidity 80% or less with no condensation |
| Storage
temperature/humidity range |
0 to
45°C, relative humidity 80% or less with no condensation |
| Interface |
RS-232C |
| Dimensions (main
unit) |
146 x
148 x 256mm |
| Weight |
4.9kg with standard
lens |
5.8kg with Small
Measuring Area Lens |
5.9kg With Small
Measuring Angle Lens |
| Standard
accessories |
Standard Lens
Macro Lens |
Standard Lens
Small Measuring Area Lens |
Standard Lens
Small Measuring Angle Lens |
| Data Processing
Software |
CS-S1w;
ND Eyepiece Filter (for finder) CS-A1; AC Adapter AC-A12;
RS-232C Cable (for IBM PC/AT, 2m, 9-pin) IF-A12 |
| Optional
accessories |
Tripod
CS-A3; Panhead CS-A4; White Calibration Plate CS-A5; ND
Filter CS-A6 (10%/for Macro Lens); ND filter CS-A7 (1%/for
Macro Lens); RS-232C Cables IF-A13 to IF-A15; Hard Case
CS-A2 (Not for Small Measuring Area Lens or Small Measuring
Angle Lens) |